AMJ-301C Mechanical Ventilator Electric Control Anesthesia Machine

AMJ-301C Mechanical Ventilator Electric Control Anesthesia Machine AMJ-301C Mechanical Ventilator Electric Control Anesthesia Machine

Applicable to closed, semi-open inhalation anesthesia, control or assisted ventilation, monitoring and display of patient ventilation parameters. Suitable for adults and children. APL valve is available, with good airtight, easy to operate, cleaning and disinfecting, concentration precise control temperature, flow, and pressure compensation techniques. Phenyl urea and sevoflurane are supplied as needed by an optional isoflurane anterior evaporator. As gas drives and electrical control, color LCD screens show various monitoring parameters.

Quantity:
Share this:
  • Product Description
  • Parameter

AMJ-301C Mechanical Ventilator Electric Control Anesthesia Machine Features:

1.Integrated breathing circuit with APL valve, excellent tightness.
2.Replacing drive bellow is NOT required when operating adult/pediatric anesthesia.
3.Applied with 1 high-precision vaporizers. (Selectatec).
4.10.1 inch screen with no dead space.
5.Real-time detection for loops, Paw, BPM, VT and compliance.
6.Comprehensive monitoring capabilities include ventilation parameters monitoring,preset parameters, alarm indication and breathing waveform monitoring.
7.ACGO available to connect with T/Bain circuit for open operation(Optional).
8.Yoke system with up to 3 back up cylinders (A type with 11L size) (Optional).

Technique Specifications

Physical Specifications

Dimensions:

70*85*140cm

Weight:

65kg

Casters:

Front wheels with lock

Screen:

10.1 inch LCD

Working Specifications

Power

AC 100V-240v, 50Hz±2%

Pipeline Supply

Gas configuration

O2, N2O, Air(Optional)

Pipeline input range

280-600kPa

Battery Power

Run-time

2180 mins

Pre-setting Functions

Language

English

Calibration

Automatic

Applications

Patient range

Audit/ Pediatric

Ventilator Specifications

Modes of Ventilation

VCV

Volume Controlled Ventilation

A/C

Assist/control Ventilation

SIMV

Synchronized Intermittent Mandatory Ventilation

MANUAL

Manual

PCV

Pressure Control Ventilation

STAND-BY

STAND-BY

Ventilator Parameter

Working type

Electronic control, pneumatic driven

Tidal Volume range

20-1500ml

Pressure range

5-60 cmbhO

Rispriation rate range

1-100 bpm (1-40bpm under SIMV)

l:E range

4:1-1:8

Ptrigger range

-20-20 cmH20 (based on PEEP)

Minute volume range

>18L/Min

SIGH range

0-5/100

System Standard

Hypoxic guard system

N2O cut-off valve, O2 concentration >25%

Safety valve

<12.5kPa

Fresh gas compensation

25-75L/min

Volume of CO2 absorber

2 L

Flow meter

4 tubes, Cascade 0.1-15L/min O2


0.1-15L/min N2O, 0.1-15L/min air(Optional)

Vaporizer

1/2(Optional) Selectetac, maximum 2 mounts(lsoflurane/Enflurane/Sevoflurane/Halothane)

Monitoring

Display graphics

Waveforms of P-T, F-T, V-T, loops of P-V,V-F,F-P

Types

VT, MV, BPM, Paw, Lung compliance, Inspiratory Platform,FiO2

Alarm Limits

Audio/Visual

No tidal volume, MV,Paw, FiO2 limitation O2,


Air failure, AC power failure, Battery low

PEEP

Type

Integrated, Electronic control

Range

OFF, 3-30cmH2O

Optional

1.AGSS

2. Suction Unit

3.Touch screen


×