Applicable to closed, semi-open inhalation anesthesia, control or assisted ventilation, monitoring and display of patient ventilation parameters. Suitable for adults and children. APL valve is available, with good airtight, easy to operate, cleaning and disinfecting, concentration precise control temperature, flow, and pressure compensation techniques. Phenyl urea and sevoflurane are supplied as needed by an optional isoflurane anterior evaporator. As gas drives and electrical control, color LCD screens show various monitoring parameters.
AMJ-301C Mechanical Ventilator Electric Control Anesthesia Machine Features:
1.Integrated breathing circuit with APL valve, excellent tightness.
2.Replacing drive bellow is NOT required when operating adult/pediatric anesthesia.
3.Applied with 1 high-precision vaporizers. (Selectatec).
4.10.1 inch screen with no dead space.
5.Real-time detection for loops, Paw, BPM, VT and compliance.
6.Comprehensive monitoring capabilities include ventilation parameters monitoring,preset parameters, alarm indication and breathing waveform monitoring.
7.ACGO available to connect with T/Bain circuit for open operation(Optional).
8.Yoke system with up to 3 back up cylinders (A type with 11L size) (Optional).
Technique Specifications | |
Physical Specifications | |
Dimensions: | 70*85*140cm |
Weight: | 65kg |
Casters: | Front wheels with lock |
Screen: | 10.1 inch LCD |
Working Specifications | |
Power | AC 100V-240v, 50Hz±2% |
Pipeline Supply | |
Gas configuration | O2, N2O, Air(Optional) |
Pipeline input range | 280-600kPa |
Battery Power | |
Run-time | 2180 mins |
Pre-setting Functions | |
Language | English |
Calibration | Automatic |
Applications | |
Patient range | Audit/ Pediatric |
Ventilator Specifications | |
Modes of Ventilation | |
VCV | Volume Controlled Ventilation |
A/C | Assist/control Ventilation |
SIMV | Synchronized Intermittent Mandatory Ventilation |
MANUAL | Manual |
PCV | Pressure Control Ventilation |
STAND-BY | STAND-BY |
Ventilator Parameter | |
Working type | Electronic control, pneumatic driven |
Tidal Volume range | 20-1500ml |
Pressure range | 5-60 cmbhO |
Rispriation rate range | 1-100 bpm (1-40bpm under SIMV) |
l:E range | 4:1-1:8 |
Ptrigger range | -20-20 cmH20 (based on PEEP) |
Minute volume range | >18L/Min |
SIGH range | 0-5/100 |
System Standard | |
Hypoxic guard system | N2O cut-off valve, O2 concentration >25% |
Safety valve | <12.5kPa |
Fresh gas compensation | 25-75L/min |
Volume of CO2 absorber | 2 L |
Flow meter | 4 tubes, Cascade 0.1-15L/min O2 |
0.1-15L/min N2O, 0.1-15L/min air(Optional) | |
Vaporizer | 1/2(Optional) Selectetac, maximum 2 mounts(lsoflurane/Enflurane/Sevoflurane/Halothane) |
Monitoring | |
Display graphics | Waveforms of P-T, F-T, V-T, loops of P-V,V-F,F-P |
Types | VT, MV, BPM, Paw, Lung compliance, Inspiratory Platform,FiO2 |
Alarm Limits | |
Audio/Visual | No tidal volume, MV,Paw, FiO2 limitation O2, |
Air failure, AC power failure, Battery low | |
PEEP | |
Type | Integrated, Electronic control |
Range | OFF, 3-30cmH2O |
Optional | |
1.AGSS | |
2. Suction Unit | |
3.Touch screen |