HB-YP90AB Hospital ICU Equipment Baby And Infant Incubator

HB-YP90AB Hospital ICU Equipment Baby And Infant Incubator HB-YP90AB Hospital ICU Equipment Baby And Infant Incubator

Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized. The heat output of these devices is usually regulated by servocontrol to keep the skin temperature constant at a site on the abdomen where a thermistor probe is attached.

Quantity:
Share this:
  • Product Description

Brief Introduction:
Two control modes: Air mode and baby mode controlled by micro-computer
Set temperature, air temperature, baby temperature are displayed separately; Natural air flow humidity
Inclination of bassinet is adjustable
Failure alarm indication;
RS-232 connector;
Oxygen input interface;
HB-90AB is configured with fixed Neonate Bilirubin Phototherapy Equipment model: HBXHZ-90 (with blue fluorescence lamp)
DC motor.

Standard Configuration:
HB-90AB: Based on HB-90A, with Neonate bilirubin phototherapy equipment model: HBXHZ-90 (fixed model).
Optional Configuration:
> 37°C temperature override mode function, Vertical height adjustment stand(VHA stand), Weighing system, SP02, Neonate bilirubin phototherapy equipment model: HBXHZ-90L (fixed model), Disposable skin temperature sensor.
Specification:
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 850VA
Mode of control: Air mode and baby mode controlled by micro-computer
Control range of air temperature: 25°C -37°C (override mode: 37-39°C is optional)
Control range of baby temperature: 34-37°C (override mode: 37-38°C is optional)
Accuracy of skin temperature sensor: ±0.2°C
The temperature difference between incubator temperature and average incubator temperature(under steady temperature condition≤0.5°C
Uniformity of temperature: 0.8°C (mattress is on the horizontal position)
Inclination of the bassinet: ±12°
Noise inside hood: 45dB(A)
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on.
The average value of total irradiance for bilirubin Ebi max of phototherapy equipment:
≥640pW/cm2 ( XHZ-90 )
≥1300pW//cm2 ( XHZ-90L )
Built-in timer range: 0-9999 hrs 59min
Accuracy of weight display: ±1% (equipped weighing system)

Requirements of environment
Environment temperature: 20 °C -30°C (Set temperature should be higher 3°C than the environment temperature)
Wind speed of environment: < 0.3m/s

Product Packaging:
Package: each unit is packed into two cartons
Dimension (Main body): 1085mm x 640mm x 910mm
(Cabinet): 1125mm x 665mm x 725mm
Gross Weight (Main body):63Kg
(Cabinet):68Kg

Baby Incubator, Transport Infant Incubator, ICU Infant Incubator, Medical Radiant Warmer, Portable Baby Warmer

×