AMJ-560B4 Hospital ICU Anesthesia Machine With Ventilator IPPV CPAP System AMJ-560B4 Hospital ICU Anesthesia Machine With Ventilator IPPV CPAP System

AMJ-560B4 Anesthesia Machine, with its ergonomic design, integrates a concise and smooth structure and high mobility.
The microcomputer-controlled Ventilator adopts exported sensors and other components.
Mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system.

Quantity:
Share this:
  • Product Description

AMJ-560B4 Hospital ICU Anesthesia Machine With Ventilator IPPV CPAP System

Product Introduction:

AMJ-560B4 Anesthesia Machine, with its ergonomic design, integrates a concise and smooth structure and high mobility. The microcomputer-controlled Ventilator adopts exported sensors and other components. Its scientific and proper design as well as rigorous production process are what demonstrate the inner quality of Aokai products: mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system. With multiple modern technologies applied, the Anesthesia Machine is both outstanding in performance and versatile in functions.


Feature:

1.  The Ventilator adopts microcomputer control with high-definition 5.7” LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.

2.  The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose two kinds of gas from Halothane, etc., ensuring a successful anesthesia operation.

3.  High-definition four-tube flowmeter, designed with stoppers for oxygen and nitrous oxide to control the output oxygen concentration.

4.  Reliable pneumatic transmission system.

5.  Economic and proper configuration of technologies, optional as required by the Customer.

6.  Modularized design and upgrading. Upgrading is available as required by the Customer.


Part of Anesthesia Machine:

Main body

High-strength engineering plastic rack, light,

beautiful and corrosion resistant

Scope of application

adult, child

Gas source

O2: 0.27~0.55Mpa

N2O: 0.27~0.55MPa

Flow meter

O2:  0.1~1.0L/min     1.1~10L/min

N20: 0.1~1.0L/min     1.1~10L/min

O2, N2O linkage and N2O Stopper

When using nitrous oxide, oxygen concentration>25%;

When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.

Flow rate of rapid oxygen supply

25~75L/min

Low oxygen pressure alarm

There will be sound alarm when the oxygen

pressure < 0.2MPa

Vaporizer

it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%.

Among Halothane, etc.,two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.

Respiratory circuit 

working mode: all-close, semi-close, semi-open

APL: 0.5~7 kPa

Respiratory Bellows

bellows for adults, bellows for children

tidal volume range:0~1,500 ml

Part of Ventilator:

Display mode

High-definition 5.7” LCD screen display

Ventilation mode

IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH

Ventilation parameters

Tidal volume:   50~1500 ml

Rate:    2~99 bpm

SIMV rate:   2~20 bpm

I:E:   2:1~1:8

Inspiratory trigger pressure:-1.0~2.0 kPa

PEEP:    0~2.0 kPa

Pressure range:    1.0~6.0 kPa

SIGH:    1.5 times the inspiratory time

Parameters for ventilation monitoring

tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure

Monitoring of oxygen concentration

(21%~100%)

Security Alarm System:

Oxygen Concentration alarm

upper limit setting range   (50%~100%)

low limit setting range    (15%~50%)

Airway pressure alarm

upper limit setting range   1.0~6.0 kPa

low limit setting range    0.4~2.0 kPa

Per-minute ventilation volume alarm

upper limit setting range   3.0~30L/min

low limit setting range    1.0~10L/min

Sustained high-pressure alarm

it will give alarm when stress have consistently

been higher than 2.5 kPa

Suffocation alarm

it will give sound and light alarm if there is

no tidal volume input for 15 seconds

Power alarm


Power  

AC 220V±10%     50Hz±1Hz

 (UPS, Storage battery)

(Mechanical Arms)           

(End-expiratory CO2 monitoring)

Items within brackets are optional.

×