HBHKN-9010 Medical Hospital Clinic Baby Care Infant Neonatal Radiant Wamer

HBHKN-9010 Medical Hospital Clinic Baby Care Infant Neonatal Radiant Wamer HBHKN-9010 Medical Hospital Clinic Baby Care Infant Neonatal Radiant Wamer

Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized. A heat shield of thin polyethylene film can be used with a radiant warmer to reduce heat loss by convection and evaporation.

Quantity:
Share this:
  • Product Description

HBHKN-9010 Infant Neonatal Radiant Wamer Introduction:
Pre-warm mode control, manual mode control and baby mode control by micro-computer;
Set temperature and baby temperature can be displayed separately;
Horizon angle of heater head?and inclination of the bassinet can be adjusted;
The panels around bassinet can be turned outward;
X-ray cassette under the bassinet;
APGAR timer;
Failure alarm indication;
RS232 connector.
Standard configuration:
Main body includes the warmer module, bassinet, main column, stand, controller, I.V. Pole and tray.
Specification:
Power supply: AC220V-230V/50Hz
Maximum Power output: 600W
Mode of control: Pre-warm mode control, manual mode control and baby mode controlled by micro-computer
Baby mode temperature control range: 34.5-37.5℃
Temperature sensor display range: 5℃-65℃
Accuracy of skin temperature sensor: ≤0.3℃
Temperature uniformity of mattress: ≤2℃
Angle of warmer module:0°,30°,60° for two ways:
Inclination of the bassinet: the three tile position;
APGAR timer: audible and visible alarm when the device runs to 50" - 1', 4' 50" - 5', 9' 50" - 10'
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on
Environmental temperature:
Operating range: 18℃~30℃
Ambient air movement rate: ≤0.3m/s

×