Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized. The heat output of these devices is usually regulated by servocontrol to keep the skin temperature constant at a site on the abdomen where a thermistor probe is attached.

Quantity:
Share this:
  • Product Description
  • Parameter

HB101/102 Infant Incubator Warmer Description:
1. Servo Control Mode: air temp;
2. Instant Parameter Displaying: set temp, air temp, and heat power rate;
3. Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air temp override, fan failure, system failure;
4. >37° Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;
5. Hood with dual-slope: 6 hand-ports; left & right cabinets, top-shelf;
6. RS-232 Output Connector (Use for network system ).
7. Export model: install the guardrail instead of top shelf

HB101

Mattress to Hood

450mm

Observation side of hood

Two incline

Operate port

6

Front access panel

826*310mm

Mattress tray size

630*360mm

Power supply

≤400VA

Temperature controlled by microcomputer

Air temp

Air temp control range

20℃-37℃

 

37℃-39℃

Skin Temp control range

-----------

 

------------

Skin Temp sensor accuracy

------------

Humidity display range

---------

Temperature fluctuation

≤0.5

Mattress temperature uniformity

≤0.8

Temperature rise time

≤30min

Noise level within hood environment

≤55dB

Alarms

 

Power failure alarm

yes

Fan failure alarm

yes

Sensor failure alarm

yes

High air temp alarm

+1.5

Low air tem alarm

-2.5

High skin temp alarm

----

Low skin tem alarm

-----

Over air tem alarm

≤38℃, ≤40

Over skin tem alarm

≤38, ≤39℃

System failure alarm

yes

Alarm silence function

yes

Operational condition

 

Environment temperature

+20℃-+30

Environment relative humidity

5%-99%

Environment air velocity of flow

0.3m/s

Others

 

RS232 interface

yes

Railing

yes

Infant bed tilt adjustment

±10° step less

Max load of mattress tray

10kg

IV Pole Max load

2kg

×